fda
3 records
FDA Recall: Lifescan Inc — Class II
20130424
When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the con | Product: LifeScan brand OneTouch¿ Select Contro...
View Source
FDA Device Recall: LifeScan brand OneTouch¿ Select Control Solution, Part Number/Model #: 02168902.
2013-04-24
Class II — When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the...
View Source
FDA Device Recall: All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Mete
2013-04-24
Class I — The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
View Source
lobbying
1 records
Lobbying Firm: LIFESCAN, INC.
Registered lobbying firm — Milpitas, CA — A global leader in blood glucose monitoring systems for home and hospital use.
View Source